Common Error Codes:
  • Er1 – possible causes:
    • sample is too small
    • problem with the test strip
    • problem with the meter
    • very low blood glucose – if the customer is suffering symptoms such as weakness, sweating, nervousness, headache or confusion they should treat themselves for hypoglycaemia.
  • Er2 – possible causes:
    • problem with the test strip
    • problem with the meter
    • very high blood glucose – if the customer is suffering from symptoms such as thirst, fatigue, excess urination or blurry vision they should follow their doctor’s recommendation for hyperglycaemia.
  • Er3 – possible causes:
    • incorrect test procedure (eg applying blood to test strip before inserting it in the meter or before the blood symbol appears)
    • problem with the test strip
    • problem with the meter
  • Er4 – possible causes:
    • problem with test strip
    • problem with meter